General |
Operating temperature
MI70
PTB330
HMP155 |
−10 … +40 °C (+14 … +104 °F)
−40 … +60 °C (−40 … +140 °F)
−80 … +60 °C (−112 … +140 °F) |
Operating humidity |
Non-condensing |
Maximum pressure limit |
5000 hPa abs. |
Power supply |
Rechargeable NiMH battery pack with
AC-adapter or 4xAA-size alkalines,
type IEC LR6 |
Menu languages |
English, Chinese, French, Spanish,
German, Russian, Japanese, Swedish,
Finnish |
Display |
LCD with backlight, graphic trend
display of any parameter, character
height up to 16 mm |
Data logging capacity |
2700 points |
Alarm |
Audible alarm function |
Compliance |
• EMC Directive (2004/108/EC)
Complies with the EMC product
family standard EN61326‑1
• Electrical equipment for
measurement control and
laboratory use
• Basic immunity test requirements
• Low Voltage Directive
(2006/95/EC)
• ROHS Directive (2002/95/EC) |
Operation Time (Using Rechargeable Battery Pack) |
Continuous use with PTB330 |
11 h typical at +20 °C (+68 °F) |
Datalogging use |
Up to 30 days |
Measurement Performance |
Barometric Pressure (PTB330) |
Measurement range |
500 … 1100 hPa |
Linearity 1) |
±0.05 hPa |
Hysteresis 1) |
±0.03 hPa |
Repeatability 1) |
±0.03 hPa |
Calibration uncertainty 2) |
±0.07 hPa |
Accuracy at +20 °C (+68 °F) 3) |
±0.10 hPa |
Temperature dependence 4) |
±0.1 hPa |
Total accuracy -40 … +60 °C (-40 … +140 °F) |
±0.15 hPa |
Long-term stability |
±0.1 hPa/year |
Settling time at startup (one sensor) |
4 s |
Response time (one sensor) |
2 s |
Acceleration sensitivity |
Negligible |
Relative Humidity (HMP155) |
Measurement range |
0 … 100 %RH |
Accuracy (incl. non-linearity, hysteresis and repeatability) |
at +15 … +25 °C (+59 … +77 °F) |
±1 %RH (0 … 90 %RH)
±1.7 %RH (90 … 100 %RH) |
at −10 … +40 °C (−4 … 104 °F) |
±(1.0 + 0.008 reading) %RH |
Factory calibration uncertainty at +20 °C (+68 °F) |
±0.6 %RH (0 … 40 %RH) 5)
±1.0 %RH (40 … 97 %RH) 5) |
Humidity sensor |
HUMICAP180R
HUMICAP180RC |
Response time at +20 °C (+68 °F) in still air with a sintered PTFE filter |
63% |
20 s |
90% |
60 s |
Temperature (HMP155) |
Measurement range |
−10 … +40 °C (+14 … +104 °F) |
Accuracy |
−10 … +20 °C (+14 … +68 °F) |
±(0.176 – 0.0028 x temperature) °C |
(+20 … +40 °C) (+68 … +104 °F) |
±(0.07 + 0.0025 x temperature) °C |
Temperature sensor |
Pt100 RTD Class F0.1 IEC 60751 |
Response time with additional temperature probe in 3 m/s air flow |
63% |
< 20 s |
90% |
< 35 s |
1) Defined as ±2 standard deviation limits of endpoint non-linearity, hysteresis, or repeatability error
2) Defined as ±2 standard deviation limits of inaccuracy of the working standard including traceability to NIST.
3) Defined as the root sum of the squares (RSS) of endpoint non-linearity, hysteresis error, repeatability error, and calibration uncertainty at room temperature.
4) Defined as ±2 standard deviation limits of temperature dependence over the operating temperature range.
5) Defined as ±2 standard deviation limits. Small variations possible, see also calibration certificate. |
Available Parameters |
Pressure parameters |
P, P3h, HCP, QFE, QNH |
Humidity and temperature parameters |
RH, T, Tdf, Td, x, Tw |
Inputs and Outputs |
MI70 probe ports |
2 |
MI70 data interface |
RS-232 (accessible only with MI70
Link software) |
PTB330 supply voltage |
10 … 35 VDC (if not powered by MI70) |
PTB330 data interface |
RS‑232C |
PTB330 serial I/O connectors |
RJ45 (service port) Male 8‑pin M12
(user port) |
HMP155 data interface |
RS‑485 |
HMP155 serial I/O connector |
Male 8‑pin M12 |
Mechanical Specifications |
PTB330 |
Housing material |
G-AlSi 10 Mg (DIN 1725) |
IP rating |
IP65 |
Pressure connector |
M5 (10-32) internal thread |
Pressure fitting |
Barbed fitting for 1/8 inch I.D. tubing
or quick connector with shutoff valve
for 1/8 inch hose |
HMP155 |
Housing material |
PC |
IP rating |
IP66 |
Additional T-probe cable length |
2 m (6 ft 6 in) |
Cable material |
PUR |
Sensor protection |
Sintered PTFE |
MI70 Measurement Indicator |
IP rating |
IP54 |
Housing material |
ABS/PC blend |
Transport Case |
IP rating (when closed) |
IP67 |
Plastic parts |
TTX01â, PP+SEBS, POM |
Metal parts |
Stainless steel AISI303 |
Interior foam material |
Polyethylene and polyether |
Weight with all instruments and typical accessories |
5.9 kg (13 lb) |
Exterior dimensions (L × W × H) |
405 × 330 × 165 mm
(15.94 × 12.99 × 6.50 in) |
Spare Parts and Accessories |
PTB330 |
MI70 – PTB330 spiral cable |
223235SP |
USB-RJ45 serial connection cable |
219685 |
Serial connection cable |
19446ZZ |
Barbed fitting 1/8 in |
19498SP |
Quick connector 1/8 in |
220186 |
Transport case with interior foams and tabletop casing for PTB330 |
Transport case with interior foams and
tabletop casing for PTB330 |
MI70 |
USB cable for MI70, includes MI70 Link software |
219687 |
MI70 Link software |
MI70LINK |
MI70 connection cable to HMT330, MMT330, DMT340, HMT100, PTB330 |
211339 |
MI70 battery pack variety of AC adapters available |
26755 |
HMP155 |
HMP155 – MI70 connection cable |
221801 |
Protection set for HMP155 calibration buttons: protective cover, 2 O-rings and protective plug |
221318 |
USB cable for HMP155 |
221040 |
Sintered teflon filter + O-ring |
219452SP |
Humidity sensor |
HUMICAP180R |
Humidity calibrator |
HMK15 |
Reviews
There are no reviews yet.